Maurice S. simplifies your size separations workflow by cutting out all the optimization guesswork. His ready-to-go CE-SDS or CE-SDS PLUS cartridges give you IgG purity data in 35 minutes. The best part? Get method development done in a day—platform methods too! Now you can run all your antibodies to get the purity data, heavy and light chain content and HC vs. NGHC composition you need.

IgG purity and heterogeneity

Maurice S. simplifies your size separations workflow by cutting out all the optimization guesswork. His ready-to-go CE-SDS or CE-SDS PLUS cartridges give you IgG purity data in 35 minutes. The best part? Get method development done in a day—platform methods too! Now you can run all your antibodies to get the purity data, heavy and light chain content and HC vs. NGHC composition you need.

More Information About Maurice S.

Feature CE-SDS Specification
Minimum Sample Volume 50 µL
Sample Delivery Electrokinetic
Typical Separation Time Reduced IgG: 25 min, Non-reduced IgG: 35 min
Detection Capability UV Absorbance at 220 nm
Typical Voltage Separation: 5750 V
Sample Injections per Cartridge 100 guaranteed, 200 maximum for CE-SDS, 500 maximum for CE-SDS PLUS
Maximum Sample Injections per Batch 48
Size Range 10–270 kDa
Sizing CV ≤2%
Relative Migration Time CV <1% for reduced IgG
Sizing Resolution ≥1.5 for NGHC/HC IgG Standard
Dynamic Range 2 logs
Linearity >0.995
Sensitivity (LOD) 0.3 µg/mL (Value based on Internal Standard)
Sample Tray Options 96-well plates or 48 vials
Dimensions 44cm H x 42cm W x 61cm D
Weight 46 kg (100 lb)