Testosterone Parameter Assay Kit Best Seller
R&D Systems, part of Bio-Techne | Catalog # KGE010
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Testosterone Parameter Assay Kit
Product Specifications
Format
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.
Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Serum
Intra-Assay Precision | Inter-Assay Precision | |||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 40 | 40 | 40 |
Mean (ng/mL) | 1.25 | 3.50 | 5.80 | 1.26 | 3.67 | 6.03 |
Standard Deviation | 0.05 | 0.11 | 0.17 | 0.07 | 0.23 | 0.41 |
CV% | 4.0 | 3.1 | 2.9 | 5.6 | 6.3 | 6.8 |
Recovery for Testosterone Parameter Assay Kit
The recovery of Testosterone spiked to levels throughout the range of the assay in various matrices was evaluated.
Sample Type | Average % Recovery | Range % |
---|---|---|
Cell Culture Media (n=4) | 103 | 95-117 |
EDTA Plasma (n=4) | 94 | 85-107 |
Heparin Plasma (n=4) | 92 | 84-101 |
Serum (n=4) | 93 | 86-104 |
Linearity
To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Testosterone were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.

Scientific Data Images for Testosterone Parameter Assay Kit
Multi-species Testosterone ELISA Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: Testosterone
Alternate Names
Additional Testosterone Products
Product Documents for Testosterone Parameter Assay Kit
Product Specific Notices for Testosterone Parameter Assay Kit
For research use only
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.