Simple Plex Human IgG Pan Specific Cartridge
ProteinSimple, part of Bio-Techne | Catalog # SPCKB-PS-007587
Customize Your Simple Plex Cartridge
Simple Plex cartridges are custom built to fit your needs. Choose from over 350 assays spanning human, mouse, rat, and non-human primate. Just select your sample type, cartridge size (16, 32, or 72 samples), and add up to 8 analytes per cartridge.
Key Product Details
Assay Length
Sample Type & Volume Required Per Well*
Sensitivity
Assay Range
Simple Plex Human IgG Pan Specific Cartridge Summary
*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay's minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.
Compatible Instruments
Ella Automated Immunoassay System
Product Specifications
Assay Type
Format
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Each control was tested multiple times in one IgG Pan Specific assay.
Inter-Assay Precision (Precision between assays) Replicates of each IgG Pan Specific control were tested in multiple assays performed by at least three technicians using two lots of reagents for human IgG Pan Specific.
EDTA Plasma, Heparin Plasma, Serum
| Intra-Assay Precision | Inter-Assay Precision | |||
|---|---|---|---|---|
| Sample | 1 | 2 | 1 | 2 |
| n | 16 | 16 | 53 | 54 |
| Mean (pg/mL) | 848 | 40,300 | 662 | 37,163 |
| Standard Deviation | 41.0 | 846 | 82.0 | 2781 |
| CV% | 4.9 | 2.1 | 12.3 | 7.5 |
Recovery for Simple Plex Human IgG Pan Specific Cartridge
Recovery was not necessary as human IgG Pan Specific shows 5 points of natural linearity.
Linearity
Samples containing and/or spiked with high concentrations of IgG Pan Specific were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay.
Scientific Data Images for Simple Plex Human IgG Pan Specific Cartridge
Data shown represents typical performance results for Lower Limit of Quantitation (LLOQ) and Upper Limit of Quantitation (ULOQ) for Simple Plex Human IgG Pan Specific Assay. Assay range is 371-90,560 pg/mL.
Preparation and Storage
Shipping
Stability & Storage
Background: IgG
R&D Systems offers a range of secondary antibodies and controls for flow cytometry, immunohistochemistry, and Western blotting. We provide species-specific secondary antibodies that are available with a variety of conjugated labels.
Our NorthernLights fluorescent secondary antibodies are bright and resistant to photobleaching. We are currently offering secondary antibodies recognizing mouse, rat, goat, sheep, and rabbit IgG as well as chicken IgY. These reagents are available with three distinct excitation and emission maxima, making them ideal for multi-color fluorescence microscopy.
Long Name
Alternate Names
Additional IgG Products
Product Documents for Simple Plex Human IgG Pan Specific Cartridge
Product Specific Notices for Simple Plex Human IgG Pan Specific Cartridge
For research use only