Simple Plex Human Follistatin Cartridge
ProteinSimple, part of Bio-Techne | Catalog # SPCKB-PS-000489
Customize Your Simple Plex Cartridge
Simple Plex cartridges are custom built to fit your needs. Choose from over 350 assays spanning human, mouse, rat, and non-human primate. Just select your sample type, cartridge size (16, 32, or 72 samples), and add up to 8 analytes per cartridge.
Key Product Details
Assay Length
Sample Type & Volume Required Per Well*
Sensitivity
Assay Range
Simple Plex Human Follistatin Cartridge Summary
*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay's minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.
Compatible Instruments
Ella Automated Immunoassay System
Product Specifications
Assay Type
Format
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Each control was tested multiple times in one Follistatin assay.
Inter-Assay Precision (Precision between assays) Replicates of each Follistatin control were tested in multiple assays performed by at least three technicians using two lots of reagents for human Follistatin.
Cell Culture Supernates, EDTA Plasma, Serum
| Intra-Assay Precision | Inter-Assay Precision | |||
|---|---|---|---|---|
| Sample | 1 | 2 | 1 | 2 |
| n | 16 | 16 | 35 | 34 |
| Mean (pg/mL) | 144 | 6411 | 149 | 6694 |
| Standard Deviation | 8.41 | 572 | 17.2 | 733 |
| CV% | 5.8 | 8.9 | 11.5 | 11.0 |
Recovery for Simple Plex Human Follistatin Cartridge
Recovery was not necessary as human Follistatin shows 5 points of natural linearity.
Linearity
Samples containing and/or spiked with high concentrations of human Follistatin were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay
Scientific Data Images for Simple Plex Human Follistatin Cartridge
Simple Plex Human Follistatin Assay Standard Curve
Data shown represents typical performance results for Lower Limit of Quantitation (LLOQ) and Upper Limit of Quantitation (ULOQ) for Simple Plex Human Follistatin Assay. Assay range is 2.77-26,410 pg/mL.Preparation and Storage
Shipping
Stability & Storage
Background: Follistatin
Follistatin (FST) is a secreted glycoprotein that binds and neutralizes several TGF-beta superfamily proteins including Activin, BMP-6, BMP-7, and GDF-8/Myostatin. It also binds Angiogenin and regulates hematopoietic stem cell adhesion to Fibronectin. Some Follistatin binding partners will also bind Follistatin-like proteins such as FSL-3. Of three follistatin isoforms, the full-length mature follistatin (FST315) is the most abundant and the sole form in plasma, but has lower binding affinity for both activins and heparins than alternative forms. The acidic tail is missing in the splice variant FST288, which shows highest affinity, and partial in the proteolytically produced FST303, which shows intermediate affinity. Genetic deletion of follistatin in mice, or expression of only the FST288 form, is lethal perinatally due to defects of lung, skin and musculoskeletal system. Expression of only the FST315 isoform allows survival, with defects in vascularization and female fertility.