Simple Plex Discovery Human Dopa Decarboxylase/DDC
ProteinSimple, part of Bio-Techne | Catalog # SPCKB-PS-015351
Customize Your Simple Plex Cartridge
Simple Plex cartridges are custom built to fit your needs. Choose from over 350 assays spanning human, mouse, rat, and non-human primate. Just select your sample type, cartridge size (16, 32, or 72 samples), and add up to 8 analytes per cartridge.
Key Product Details
Assay Length
Sample Type & Volume Required Per Well*
Sensitivity
Assay Range
Simple Plex Discovery Human Dopa Decarboxylase/DDC Summary
*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay's minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.
Compatible Instruments
Ella Automated Immunoassay System
Product Specifications
Assay Type
Format
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Each control was tested multiple times in one DDC assay.
Inter-Assay Precision (Precision between assays) Replicates of each DDC control were tested in multiple assays performed by at least three technicians using two lots of reagents for human DDC.
Cell Culture Supernates, Cerebrospinal Fluid, EDTA Plasma, Heparin Plasma, Serum
| Intra-Assay Precision | Inter-Assay Precision | |||
|---|---|---|---|---|
| Sample | 1 | 2 | 1 | 2 |
| n | 16 | 16 | 35 | 35 |
| Mean (pg/mL) | 19.8 | 916 | 18.5 | 858 |
| Standard Deviation | 0.494 | 28.3 | 1.73 | 56.9 |
| CV% | 2.5 | 3.1 | 9.3 | 6.6 |
Recovery for Simple Plex Discovery Human Dopa Decarboxylase/DDC
Recovery for the human DDC assays was evaluated at three different spiked concentrations of DDC protein.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=2) | 103 | 102-103 |
| Cerebrospinal Fluid (n=2) | 105 | 104-105 |
Linearity
Samples containing and/or spiked with high concentrations of DDC were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay.
Scientific Data Images for Simple Plex Discovery Human Dopa Decarboxylase/DDC
Data shown represents typical performance results for Lower Limit of Quantitation (LLOQ) and Upper Limit of Quantitation (ULOQ) for Simple Plex Discovery Human DDC Assay. Assay range is 0.63-6000 pg/mL.
Preparation and Storage
Shipping
Stability & Storage
Background: Dopa Decarboxylase/DDC
Additional Dopa Decarboxylase/DDC Products
Product Documents for Simple Plex Discovery Human Dopa Decarboxylase/DDC
Product Specific Notices for Simple Plex Discovery Human Dopa Decarboxylase/DDC
For research use only