Simple Plex AAV8 Cartridge Best Seller
ProteinSimple, part of Bio-Techne | Catalog # SPCKB-OT-007970
Need a different panel or sample type? Customize your cartridge here.
Key Product Details
Assay Length
Sample Type & Volume Required Per Well*
Sensitivity
0.1 106 capsids/mL (Bioprocess Samples)
Assay Range
2.47-9428 106 capsids/mL (Bioprocess Samples)
Simple Plex AAV8 Cartridge Summary
*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay's minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.
Compatible Instruments
Ella Automated Immunoassay System
Product Specifications
Assay Type
Format
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Each control was tested 16 times in one assay.
Inter-Assay Precision (Precision between assays) Replicates of each control were tested in multiple assays performed by at least three technicians using two lots of reagents.
Bioprocess Samples
| Intra-Assay Precision | Inter-Assay Precision | |||
|---|---|---|---|---|
| Sample | 1 | 2 | 1 | 2 |
| n | 16 | 16 | 47 | 45 |
| Mean (10<sup>6</sup> capsids/mL) | 23.1 | 1073 | 49.2 | 2202 |
| Standard Deviation | 0.658 | 55.4 | 4.51 | 99.3 |
| CV% | 2.9 | 5.2 | 9.2 | 4.5 |
Recovery for Simple Plex AAV8 Cartridge
Recovery was not necessary as AAV8 shows natural linearity data from a neat sample to a 1:16 dilution.
Linearity
Samples containing and/or spiked with high concentrations of AAV8 were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay.
Scientific Data Images for Simple Plex AAV8 Cartridge
AAV8 Simple Plex Assay Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: AAV8
Long Name
Additional AAV8 Products
Product Documents for Simple Plex AAV8 Cartridge
Product Specific Notices for Simple Plex AAV8 Cartridge
For research use only