Human Erythropoietin/EPO Quantikine ELISA Kit, RUO
R&D Systems, part of Bio-Techne | Catalog # DEPRU0B
Key Product Details
Assay Length
Sample Type & Volume Required Per Well
Sensitivity
Assay Range
Product Summary for Human Erythropoietin/EPO Quantikine ELISA Kit, RUO
Product Specifications
Measurement
Detection Method
Conjugate
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess Intra-Assay Precision.
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in twenty separate assays to assess Inter-Assay Precision. Assays were performed by at least three technicians using two lots of components.
EDTA Plasma, Serum
| Intra-Assay Precision | Inter-Assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean (mIU/mL) | 10.3 | 40.3 | 117 | 10.4 | 41.9 | 118 |
| Standard Deviation | 0.370 | 1.45 | 1.97 | 0.860 | 2.20 | 5.78 |
| CV% | 3.6 | 3.6 | 1.7 | 8.3 | 5.3 | 4.9 |
Recovery for Human Erythropoietin/EPO Quantikine ELISA Kit, RUO
The recovery of human Epo spiked to levels throughout the range of the assay in various matrices was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| EDTA Plasma (n=4) | 94 | 84-98 |
| Serum (n=4) | 101 | 96-107 |
Linearity
To assess the linearity of the assay, samples spiked with high concentrations of human Epo in various matrices were diluted with calibrator diluent to produce samples with values within the dynamic range of the assay.
Scientific Data Images for Human Erythropoietin/EPO Quantikine ELISA Kit, RUO
Human Erythropoietin Quantikine ELISA Standard Curve
Preparation and Storage
Shipping
Stability & Storage
Background: Erythropoietin/EPO
Erythropoietin/EPO is a renal glycoprotein hormone that promotes erythrocyte formation by preventing the apoptosis of early erythroid precursors which express the EPO receptor (EPO R). Rapid regulation of circulating EPO allows tight control of erythrocyte production and hemoglobin concentrations. Anemia or other causes of low tissue oxygen tension induce EPO production by stabilizing the hypoxia-induceable transcription factors HIF-1 alpha and HIF-2 alpha. Erythropoietin/EPO additionally plays a tissue-protective role in ischemia by blocking apoptosis and inducing angiogenesis.
R&D System products include Erythropoietin antibodies, proteins, and ELISAs. Our products cover multiple species and our EPO antibodies are validated for multiple applications including Neutralization, Western blot, ELISA, and Immunohistochemistry.