Simple Plex Human D-dimer Cartridge
ProteinSimple, part of Bio-Techne | Catalog # SPCKB-PS-001976
Customize Your Simple Plex Cartridge
Simple Plex cartridges are custom built to fit your needs. Choose from over 350 assays spanning human, mouse, rat, and non-human primate. Just select your sample type, cartridge size (16, 32, or 72 samples), and add up to 8 analytes per cartridge.
Key Product Details
Assay Length
Sample Type & Volume Required Per Well*
Sensitivity
Assay Range
Simple Plex Human D-dimer Cartridge Summary
*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay's minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.
Compatible Instruments
Ella Automated Immunoassay System
Product Specifications
Assay Type
Format
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Each control was tested multiple times in one D-Dimer assay.
Inter-Assay Precision (Precision between assays) Replicates of each D-Dimer control were tested in multiple assays performed by at least three technicians using two lots of reagents for human D-Dimer.
Cell Culture Supernates, Citrate Plasma, EDTA Plasma, Heparin Plasma, Serum
| Intra-Assay Precision | Inter-Assay Precision | |||
|---|---|---|---|---|
| Sample | 1 | 2 | 1 | 2 |
| n | 16 | 16 | 34 | 34 |
| Mean (pg/mL) | 3288 | 174,738 | 3282 | 173,674 |
| Standard Deviation | 226 | 8198 | 267 | 11,649 |
| CV% | 6.9 | 4.7 | 8.1 | 6.7 |
Recovery for Simple Plex Human D-dimer Cartridge
Recovery was not necessary as human D-Dimer shows 5 points of natural linearity.
Linearity
Samples containing and/or spiked with high concentrations of human D-Dimer were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay
Scientific Data Images for Simple Plex Human D-dimer Cartridge
Simple Plex Human D-Dimer Assay Standard Curve
Data shown represents typical performance results for Lower Limit of Quantitation (LLOQ) and Upper Limit of Quantitation (ULOQ) for Simple Plex Human D-Dimer Assay. Assay range is 142-540,890 pg/mL.Preparation and Storage