Simple Plex Human alpha-Synuclein Cartridge
ProteinSimple, part of Bio-Techne | Catalog # SPCKC-PS-011989
Customize Your Simple Plex Cartridge
Simple Plex cartridges are custom built to fit your needs. Choose from over 350 assays spanning human, mouse, rat, and non-human primate. Just select your sample type, cartridge size (16, 32, or 72 samples), and add up to 8 analytes per cartridge.
Key Product Details
Assay Length
Sample Type & Volume Required Per Well*
Sensitivity
Assay Range
Simple Plex Human alpha-Synuclein Cartridge Summary
*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay's minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.
Compatible Instruments
Ella Automated Immunoassay System
Product Specifications
Reactivity
Specificity
Cross-reactivity
Interference
Precision
Intra-Assay Precision (Precision within an assay) Each control was tested 16 times in one assay.
Inter-Assay Precision (Precision between assays) Replicates of each control were tested in multiple assays performed by at least three technicians using two lots of reagents.
Cell Culture Supernates, Cerebrospinal Fluid, EDTA Plasma, Heparin Plasma, Serum
| Intra-Assay Precision | Inter-Assay Precision | |||
|---|---|---|---|---|
| Sample | 1 | 2 | 1 | 2 |
| n | 16 | 16 | 54 | 54 |
| Mean (pg/mL) | 22.2 | 862 | 16.6 | 745 |
| Standard Deviation | 0.974 | 60.5 | 1.45 | 79.2 |
| CV% | 4.4 | 7.0 | 8.8 | 10.6 |
Recovery for Simple Plex Human alpha-Synuclein Cartridge
Recovery at three different spiked concentrations within the range of the assay was evaluated.
| Sample Type | Average % Recovery | Range % |
|---|---|---|
| Cell Culture Media (n=2) | 115 | 115-116 |
| Cerebrospinal Fluid (n=4) | 115 | 109-118 |
Linearity
Samples containing and/or spiked with high concentrations of recombinant human alpha-Synuclein were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay.
Scientific Data Images for Simple Plex Human alpha-Synuclein Cartridge
Human alpha-Synuclein Ella Assay Standard Curve
Human alpha-Synuclein Ella Assay Endogenous Levels
Functional LOD and functional LLOQ are calculated as the LOD and LLOQ of the assay multiplied by the minimum required dilution (MRD) for the sample type.Preparation and Storage