Skip to main content

Simple Plex HIV-1 Gag p24 Cartridge

ProteinSimple, part of Bio-Techne | Catalog # SPCKB-PS-000981

This product has been discontinued and replaced with the HIVp24 2nd generation Simple Plex Assay.
Discontinued Product
SPCKB-PS-000981 has been discontinued. View all HIV-1 Gag p24 products.

Key Product Details

Assay Length

90 minutes

Sample Type & Volume Required Per Well*

Cell Culture Supernates (25 µL), Serum (25 µL), EDTA Plasma (25 µL), Heparin Plasma (25 µL)

Sensitivity

0.67 pg/mL

Assay Range

3.93-15000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma)

Simple Plex HIV-1 Gag p24 Cartridge Summary

Simple Plex assay for the detection of HIV Gag p24 in cell culture supernatant (CCS), serum, and plasma (EDTA/Heparin) on Ella automated benchtop ELISA. 

*The Sample Volume represented is based on the amount of sample incorporated into the reaction after taking into account the assay's minimum required dilution for a given matrix. Serial dilution may be necessary to achieve some of the final sample volumes represented.

Compatible Instruments

Ella Automated Immunoassay System

Ella Automated Immunoassay System

Efficiently analyze biomarkers with Ella's advanced technology, providing accurate, precise results in under 90 minutes. Quantify up to 8 targets per assay through a streamlined automated workflow. Choose from >350 highly validated Simple Plex assays or Simple Plex discovery assays.

Product Specifications

Assay Type

Automated sandwich ELISA

Format

Microfluidic cartridge containing three Glass Nano Reactors (GNR)

Specificity

Natural and recombinant HIV-1 Gag p24.

Cross-reactivity

< 0.5% cross-reactivity observed with available related molecules. < 50% cross-species reactivity observed with species tested.

Interference

No significant interference observed with available related molecules.

Precision

Intra-Assay Precision (Precision within an assay) Each control was tested 16 times in one assay

Inter-Assay Precision (Precision between assays) Replicates of each control were tested in multiple assays performed by at least three technicians using two lots of reagents

Cell Culture Supernates, EDTA Plasma, Heparin Plasma, Serum

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 1 2
n 16 16 35 34
Mean (pg/mL) 105 4500 94.3 4074
Standard Deviation .37 237 9.64 377
CV% 7.1 5.3 10.2 9.3

Recovery for Simple Plex HIV-1 Gag p24 Cartridge

Recovery at three different spiked concentrations within the range of the assay was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Supernates (n=4) 114 111-118
EDTA Plasma (n=4) 107 102-116
Heparin Plasma (n=4) 107 99-112
Serum (n=4) 111 109-113

Linearity

Samples containing and/or spiked with high concentrations of human HIV Gag p24 were serially diluted with Sample Diluent to produce samples within the dynamic range of the assay

HIV Gag p24 Assay Linearity

Scientific Data Images for Simple Plex HIV-1 Gag p24 Cartridge

HIV Gag p24 Assay Standard Curve

HIV Gag p24 Assay Standard Curve

Preparation and Storage

Shipping

The product is shipped with polar packs. Upon receipt, store it immediately at the temperature recommended below.

Stability & Storage

Store the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: HIV-1 Gag p24

The gag gene of human immunodeficiency virus 1 (HIV-1) encodes a precursor protein known as Pr55Gag. The viral protease PR cleaves this precursor to generate p17, p24, p7, and p6 proteins which are required for virus particle assembly. p24 is a major viral core structural protein. Its measurement is commonly used as an indicator of HIV-1 infection and viral load.

Long Name

Human Immunodeficiency Virus Type 1 GAG/Group-specific Antigen core protein of 24 kDa

Alternate Names

CA, Capsid Protein, HIV1 Gag p24, Pr55(Gag)

Entrez Gene IDs

155030 (Viral)

Gene Symbol

GAG

Additional HIV-1 Gag p24 Products

Product Documents for Simple Plex HIV-1 Gag p24 Cartridge

Loading product document information...

Loading product document information...

Certificate of Analysis

To download a Certificate of Analysis, please enter your Simple Plex catalog number and lot number in the search boxes below.

Product Specific Notices for Simple Plex HIV-1 Gag p24 Cartridge

For research use only

Loading...